HYTAIS TS600

HYTAIS TS600 Infrared Non-Contact Thermometer User Manual

ماډل: TS600

نښه: هایټیس

1. پیژندنه

The HYTAIS TS600 is a non-contact infrared thermometer designed for quick and accurate surface temperature measurements. It offers a wide temperature range, fast response time, and adjustable emissivity, making it suitable for various applications from home use to industrial inspections.

This device allows for safe temperature detection without direct contact, which is ideal for measuring high-temperature equipment, hazardous chemicals, or hard-to-reach objects.

د خوندیتوب معلومات

  • لیزر په مستقیم ډول سترګو ته یا په غیر مستقیم ډول د انعکاس سطحو څخه مه مه نیسئ.
  • Do not use the thermometer near explosive gases, steam, or dust.
  • Keep the device away from electromagnetic fields (e.g., arc welders, induction heaters).
  • واحد په اوبو کې مه ډوبوئ.
  • This product is suitable for temperature measurement on the surface of opaque objects.
  • When measuring, keep the distance from the target within 30cm for optimal accuracy.
  • Avoid strong reflection environments, glass, liquid, and other transparent materials, as these can affect measurement accuracy. For such materials, specialized temperature measurement equipment is recommended.

2. کلیدي ځانګړتیاوې

HYTAIS TS600 Infrared Thermometer highlighting key features

Figure 2.1: HYTAIS TS600 with highlighted features

  • د غیر تماس اندازه کول: Safely measures surface temperatures without direct contact.
  • د تودوخې پراخه لړۍ: اندازه یې له -50℃ څخه تر 600℃ (-58℉ څخه تر 1112℉) پورې ده.
  • چټک غبرګون: Provides readings in approximately 0.5 seconds.
  • د تنظیم وړ اخراج: Emissivity adjustable from 0.1 to 1.0 to adapt to various surface types for improved accuracy.
  • د تودوخې دوه ګوني واحدونه: په اسانۍ سره د سیلسیس (℃) او فارنهایټ (℉) ترمنځ بدل کړئ.
  • د لیزر موقعیت: Equipped with an auxiliary laser point for accurate targeting of the measurement area.
  • HD Backlit Screen: Clear digital display with backlight for readability in various lighting conditions.
  • اتوماتیک بندول: د غیر فعالیت دورې وروسته په اتوماتيک ډول بندیدو سره د بیټرۍ ژوند ساتي.
  • د ټیټ بیټرۍ شاخص: کله چې بیټرۍ بدلولو ته اړتیا ولري، کارونکي ته خبرداری ورکوي.
  • د لوړ / ټیټ حرارت الارم: User-settable alarms for temperatures exceeding or falling below specified thresholds.
Additional functions of the thermometer including backlight, auto shut-off, C/F selection, and low battery indicator

Figure 2.2: Additional functions of the HYTAIS TS600

3. د محصول اجزا

Diagram showing the labeled parts of the HYTAIS TS600 thermometer: Laser, Thermometer probe, HD screen, MODE button, Laser aiming on/off button, °C/°F convert button, Power button, Battery holder.

شکل ۳: د پینل پیژندنه او اجزا

  1. لیزر: Emits a laser dot for precise targeting of the measurement area.
  2. د ترمامیتر تحقیقات: The infrared sensor for temperature detection.
  3. HD سکرین: Digital display for temperature readings and settings.
  4. د موډ تڼۍ: Used to access and adjust settings like emissivity and alarm thresholds.
  5. Laser Aiming On/Off Button: Toggles the laser pointer.
  6. °C/°F Convert Button: د سیلسیس او فارنهایټ د تودوخې واحدونو ترمنځ بدلیږي.
  7. Power Button (Trigger): Activates the thermometer and initiates measurement.
  8. د بیټرۍ لرونکی: د AAA بیټرۍ لپاره خونه.

4. ترتیب کول

4.1 د بیټرۍ نصب کول

The HYTAIS TS600 requires two (2) 1.5V AAA batteries (not included) for operation. Follow these steps to install the batteries:

  1. Locate the battery holder compartment at the bottom of the handle.
  2. د بیټرۍ کمپارټ پوښ خلاص کړئ.
  3. Insert two 1.5V AAA batteries, ensuring correct polarity (+ and -).
  4. د بیټرۍ کمپارټ پوښ په خوندي ډول وتړئ.
Diagram showing easy battery replacement for the thermometer, indicating 2x1.5V AAA batteries are not included.

شکل 4.1: د بیټرۍ نصب کول

5. عملیات

۵.۲ د تودوخې اساسي اندازه کول

  1. ترمامیتر په نښه شوي شی باندې وښایاست.
  2. Pull and hold the Power Button (trigger) to activate the device and begin measurement. The temperature reading will appear on the HD screen.
  3. په ښودنه کې د وروستي اندازه شوي تودوخې ساتلو لپاره محرک خوشې کړئ.

5.2 Laser Aiming

The built-in laser helps to accurately target the measurement area.

  • To turn on/off the laser: Short press the "Laser Aiming On/Off" button (refer to Figure 3.1 for button location).
Image showing the thermometer with laser aiming at cupcakes in an oven, indicating wide range and fast reading.

Figure 5.1: Laser Aiming for precise measurement

5.3 Switching Temperature Units (℃/℉)

د سیلسیس (℃) او فارنهایټ (℉) ترمنځ د بدلولو لپاره:

  • Short press the "°C/°F Convert" button (refer to Figure 3.1 for button location). The display will change accordingly.
Image showing the thermometer display switching between Celsius and Fahrenheit.

Figure 5.2: Temperature Unit Switching

۴.۵ د خپریدو تنظیم

Emissivity (ε) is a measure of an object's ability to emit infrared energy. Different materials have different emissivities. Adjusting the emissivity setting on your thermometer ensures more accurate readings for various surfaces. The TS600 allows adjustment from 0.1 to 1.0.

  1. Long press the "MODE" button to enter the emissivity setting. The current emissivity value (e.g., "ε=0.95") will flash on the screen.
  2. د اخراج ارزښت تنظیمولو لپاره د پورته (▲) او ښکته (▼) تڼیو څخه کار واخلئ.
  3. Press the "MODE" button again to confirm the setting and exit.

Common Emissivity Values:

  • Cement: 0.96
  • Oil: 0.94
  • Wood: 0.94
  • Most organic materials, painted or oxidized surfaces: 0.95 (default)
Image showing adjustable emissivity for different materials like cement, oil, and wood.

Figure 5.3: Adjustable Emissivity for High Accuracy

5.5 High / Low Temperature Alarm Setting

The thermometer can be set to alert you if the measured temperature exceeds a high threshold or falls below a low threshold.

  1. Long press the "MODE" button until "Hi" appears on the screen. This is the High Temperature Alarm setting.
  2. Use the Up (▲) and Down (▼) buttons to set the desired high temperature threshold.
  3. Press the "MODE" button again to switch to the Low Temperature Alarm setting ("Low" will appear).
  4. Use the Up (▲) and Down (▼) buttons to set the desired low temperature threshold.
  5. Press the "MODE" button one more time to exit the alarm settings.

5.6 Distance to Spot Ratio (D:S)

The D:S ratio for the HYTAIS TS600 is 12:1. This means that for every 12 units of distance from the target, the measurement spot diameter is 1 unit. For example, at a distance of 12 inches, the measurement spot will be 1 inch in diameter.

To ensure accurate readings, the target object should be larger than the measurement spot. As the distance to the object increases, the measurement spot size also increases. It is recommended that the target being measured is at least twice the diameter of the thermometer's spot.

Diagram explaining the 12:1 Distance Spot Ratio (D:S) for the infrared thermometer.

Figure 5.4: Understanding Distance Spot Ratio

6. مشخصات

Product standard image showing the thermometer and its packaging.

Figure 6.1: Product Standard View

صفتارزښت
د ماډل شمیرهTS600
نظریهڅو اړخیزه ترمامیتر
د حرارت درجه اندازه کول-50℃~600℃ (-58℉-1112℉)
د تودوخې اعظمي اندازه کول۱۲۰ درجې سانتي ګراد او پورته
د ځای نسبت ته واټن (D:S)12:1
ایمسیوتیتAdjustable 0.1-1.0 (Default 0.95)
طول موج630-670nm
د غبرګون وخت0.5 ثانیې
Power Type/Source2 x 1.5V AAA Battery (not included)
د ښودلو اندازه۲.۹ انچه او لاندې
د ښودلو ډولډیجیټل (LCD)
سټایللاسي
کارولصنعتي
تصدیقCE، RoHS
د محصول ابعاد (تقریبا)Length: 14.2 cm (5.6 in), Width: 8.6 cm (3.4 in), Depth: 4.5 cm (1.8 in)
بیټرۍ پکې شامله دهنه
لوړ اندیښمن کیمیاوي موادهیڅ نه
سمارټ وسیله دهنه
Diagram showing the product dimensions in centimeters and inches.

شکل ۲: د محصول ابعاد

7. ستونزه حل کول

ستونزهاحتمالي لاملد حل لاره
وسیله نه ورځي.مړه یا په غلط ډول نصب شوي بیټرۍ.Check battery polarity and replace with fresh 1.5V AAA batteries.
د تودوخې ناسم لوستل.Incorrect emissivity setting; target too small or too far; reflective surface; transparent material being measured.Adjust emissivity for the specific material. Ensure target is larger than the measurement spot and within optimal distance (D:S ratio). Avoid measuring through glass or liquids.
Laser not working.Laser is turned off.Short press the "Laser Aiming On/Off" button to activate the laser.
ښودنه تیاره یا ځلیدونکی ده.د بیټرۍ ټیټ ځواک.بیټرۍ بدل کړئ.

8. ساتنه

8.1 پاکول

  • لینز: Use a soft, clean cloth or cotton swab with water or medical alcohol. Wipe gently.
  • Casing: Clean the instrument casinد اعلان سره gamp ټوکر او نرم صابون. د کثافاتو پاکوونکي یا محلول مه کاروئ.
  • Ensure the lens is clean and free of dust or debris for accurate measurements.

8.2 ذخیره کول

  • ترمامیتر په یخ او وچ ځای کې وساتئ کله چې نه کارول کیږي.
  • که چیرې وسیله د اوږدې مودې لپاره د لیکیدو مخنیوي لپاره ونه کارول شي، نو بیټرۍ لرې کړئ.
  • د ماشومانو د لاسرسي څخه لرې وساتئ.

9. د کارن لارښوونې

  • For best accuracy, always ensure the target object is larger than the measurement spot.
  • When measuring liquids, measure the surface of the liquid, not through the liquid itself.
  • Be aware of highly reflective surfaces, as they can cause inaccurate readings. Adjust emissivity or use a non-reflective tape if possible.
  • The thermometer is designed for surface temperature measurement; it is not intended for measuring internal temperatures of objects or body temperature.

10. تضمین او ملاتړ

The HYTAIS TS600 Infrared Non-Contact Thermometer is designed for reliability and durability. For any questions, technical support, or warranty claims, please refer to the contact information provided by your retailer or the manufacturer's official webسایټ

مهرباني وکړئ د تضمین موخو لپاره د پیرود ثبوت وساتئ.

۱۰. د محصول د ښودلو ویډیو

Video 11.1: Unboxing and Basic Operation of the HYTAIS TS600 Infrared Thermometer. This video demonstrates how to unbox the thermometer, install batteries, turn the laser on/off, switch between Celsius and Fahrenheit, adjust emissivity, and set high/low temperature alarms.


د HYTAIS TS600 غیر تماس انفراریډ ترمامیتر: ځانګړتیاوې او عملیاتي لارښود

د HYTAIS TS600 غیر تماس انفراریډ ترمامیتر: ځانګړتیاوې او عملیاتي لارښود

۰:۴۵ • ۱۹۲۰×۱۰۷۸ • فیچر_ډیمو
د HYTAIS TS600 ډیجیټل انفراریډ ترمامیتر د ان باکس کولو او ځانګړتیا ښودلو څرنګوالی

د HYTAIS TS600 ډیجیټل انفراریډ ترمامیتر د ان باکس کولو او ځانګړتیا ښودلو څرنګوالی

۰:۴۵ • ۱۹۲۰×۱۰۷۸ • فیچر_ډیمو

اړوند اسناد - TS600

مخکېview د HYTAIS TS380/TS600 انفراریډ ترمامیتر کارونکي لارښود
د HYTAIS TS380 او TS600 غیر تماس لرونکي انفراریډ ترمامیترونو لپاره د کارونکي لارښود، د سطحې د تودوخې دقیق اندازه کولو لپاره د ځانګړتیاوو، خوندیتوب لارښوونو، عملیاتو، ساتنې او مشخصاتو تفصیل.
مخکېview د HYTAIS TS600 ډیجیټل انفراریډ ترمامیتر ټوپک کارونکي لارښود
د HYTAIS TS600 ډیجیټل انفراریډ ترمامیتر ټوپک لپاره د کارونکي لارښود، د هغې ځانګړتیاوې، عملیات، د خوندیتوب احتیاطي تدابیر، ساتنه، او تضمین معلومات په تفصیل سره بیانوي.
مخکېview د HYTAIS TS-U7 په بشپړ ډول اتوماتیک پیژندنه ملټي میټر کارونکي لارښود
د HYTAIS TS-U7 بشپړ اتوماتیک پیژندنې ملټي میټر لپاره جامع کارونکي لارښود. د دقیق بریښنایی اندازه کولو لپاره د هغې ځانګړتیاو، مشخصاتو، عملیاتو او خوندیتوب لارښوونو په اړه زده کړه وکړئ.
مخکېview د HYTAIS TS20A قلم ډوله ډیجیټل ملټي میټر کارونکي لارښود
د HYTAIS TS20A قلم ډوله ډیجیټل ملټي میټر لپاره د کارونکي لارښود، د هغې ځانګړتیاوې، عملیات، د خوندیتوب احتیاطي تدابیر، او د AC/DC حجم اندازه کولو لپاره مشخصات په تفصیل سره بیانوي.tage، مقاومت، او دوام.