GIMA 28211

د GIMA 28211 سوپر ویګا سکشن اسپریټر کارونکي لارښود

Model: 28211 | Brand: GIMA

1. پیژندنه

The GIMA 28211 Super Vega Suction Aspirator is a compact and reliable medical device designed for the aspiration of bodily fluids such as mucus, catarrh, or blood. It is suitable for nasal, oral, and tracheal aspiration in both adults and children. This manual provides essential information for the safe and effective operation, maintenance, and troubleshooting of your device.

GIMA 28211 Super Vega Suction Aspirator main unit

Image 1.1: The GIMA 28211 Super Vega Suction Aspirator, showing the main unit with collection jar, tubing, and control panel.

2. د خوندیتوب معلومات

  • د وسیلې له چلولو دمخه دا لارښود په بشپړ ډول ولولئ.
  • Ensure the device is placed on a stable, flat surface during operation.
  • اصلي واحد په اوبو یا کوم بل مایع کې مه ډوبوئ.
  • Only use accessories and spare parts approved by GIMA.
  • Regularly check the power cord for damage. Do not use if damaged.
  • وسیله د ماشومانو او غیر مجاز پرسونل له لاسرسي لرې وساتئ.
  • This device is intended for medical use and should be operated by trained individuals or under medical supervision.
  • د عملیاتو په جریان کې د وسیلې شاوخوا مناسب هوا ورکول ډاډمن کړئ.

3. د بسته محتويات

ډاډ ترلاسه کړئ چې ټولې برخې شتون لري او د خلاصولو پر مهال زیانمن شوي نه دي:

  • GIMA 28211 Super Vega Suction Aspirator main unit
  • 1000 ml autoclavable collection jar with lid
  • Suction tubing set
  • د باکتریا فلټر
  • د بریښنا تار
  • د کارن لارښود
Collection jar and lid

Image 3.1: 1000 ml collection jar with lid.

د باکتریا فلټر

Image 3.2: Bacterial filter.

د بریښنا تار

Image 3.3: Power cord.

4. ترتیب کول

  1. ځای په ځای کول: Place the aspirator on a stable, flat, and dry surface, ensuring adequate space for ventilation.
  2. Connect Collection Jar: Securely attach the 1000 ml collection jar to the designated holder on the side of the main unit. Ensure the lid is tightly sealed.
  3. ټیوب نښلول:
    • Connect one end of the short silicone tube to the 'VACUUM' port on the collection jar lid.
    • Connect the other end of this short tube to the suction inlet on the main unit (usually marked 'SUCTION' or similar).
    • Connect one end of the longer silicone tube to the 'PATIENT' port on the collection jar lid. This end will connect to the patient catheter.
  4. Install Bacterial Filter: Insert the bacterial filter between the 'VACUUM' port on the collection jar lid and the suction inlet of the main unit. This prevents fluid overflow and protects the pump. Ensure the filter is correctly oriented according to any arrows on its casing.
  5. د بریښنا پیوستون: Connect the power cord to the aspirator's power inlet and then plug it into a suitable electrical outlet (220-230 V/50-60 Hz).
GIMA 28211 Super Vega Suction Aspirator fully assembled

Image 4.1: The aspirator fully assembled and ready for operation.

5. عملیاتي لارښوونې

  1. پاور آن: Press the green power switch (marked with 'I/O') located on the front panel to turn on the device. The indicator light should illuminate.
  2. سکشن تنظیم کړئ: Use the 'SUCTION CONTROL' knob on the front panel to adjust the desired vacuum level. Turn clockwise to increase suction and counter-clockwise to decrease it.
  3. د څارنې ویکیوم: Observe the vacuum gauge on the front panel to ensure the desired suction pressure is achieved. The gauge displays pressure in mmHg and kPa.
  4. کتیټر وصل کړئ: Attach the appropriate sterile suction catheter to the free end of the patient tubing.
  5. Perform Aspiration: Carefully perform the aspiration procedure as directed by a medical professional.
  6. بندول: After use, press the power switch to turn off the device.
Control panel of GIMA 28211 Super Vega Suction Aspirator

Image 5.1: Close-up of the control panel, showing the power switch, vacuum gauge, and suction control knob.

6. ساتنه او پاکول

Regular cleaning and maintenance are crucial for the device's longevity and hygienic operation.

6.1. Collection Jar

  • Empty the collection jar after each use or when the fluid level approaches the maximum fill line.
  • The 1000 ml collection jar is autoclavable at 121°C. An optional jar is available that can be autoclaved at 134°C.
  • Wash the jar and lid thoroughly with warm water and a mild detergent. Rinse well and sterilize as appropriate.
  • Ensure the safety float valve in the lid is clean and moves freely to prevent overflow.

6.2. Tubing and Filter

  • Clean the suction tubing after each use by aspirating clean water through it.
  • Replace the bacterial filter regularly, especially if it becomes wet, discolored, or shows signs of contamination. The filter is disposable.

6.3. اصلي واحد

  • Wipe the exterior of the main unit with a soft cloth dampened with a mild disinfectant.
  • د کثافاتو پاکوونکي یا محلول مه کاروئ.
  • Ensure no liquids enter the internal components of the device.

7. ستونزه حل کول

ستونزهاحتمالي لاملد حل لاره
وسیله فعاله نه دهNo power supply; damaged power cord; faulty power switchCheck power cord connection and wall outlet. Inspect power cord for damage. If problem persists, contact technical support.
نه یا کمزوری سکشنLoose connections; clogged tubing/catheter; full collection jar; faulty bacterial filter; pump issueCheck all tubing connections for leaks. Clear any blockages in tubing/catheter. Empty collection jar. Replace bacterial filter. If problem persists, contact technical support.
Fluid overflow into main unitCollection jar overfilled; faulty safety float valve; bacterial filter not installed or cloggedEmpty collection jar immediately. Check float valve for proper movement. Ensure bacterial filter is correctly installed and not clogged.
د عملیاتو پرمهال غیر معمولي شورLoose components; internal pump issueEnsure all external components are secure. If noise persists, discontinue use and contact technical support.

8. مشخصات

  • ماډل: GIMA 28211 Super Vega
  • د عملیاتو Voltage: 220-230V / 50-60Hz
  • د بریښنا مصرف: 106 W
  • Collection Jar Capacity: 1000 ملی لیتر
  • ابعاد (L x W x H): 35 x 21 x 18 سانتي متره
  • وزن: تقریبا 3.34 کیلوګرامه
  • د ویکیوم تنظیم کوونکی: په دوامداره توګه تنظیم کیدونکی
  • Autoclavable Jar: 121°C (optional 134°C)
  • موصلیت: Electrical and thermal insulating plastic body

9. تضمین او ملاتړ

Information regarding the specific warranty period for the GIMA 28211 Super Vega Suction Aspirator is not provided in this document. Please refer to your purchase documentation or contact your local GIMA distributor or customer service for warranty details and technical support.

For assistance, please have your model number (28211) and purchase date available.

اړوند اسناد - 28211

مخکېview د ګیما سوپر ویګا بیټرۍ سکشن یونټ د کارونکي لارښود او تخنیکي لارښود
دا سند د GIMA SUPERVEGA بیټرۍ سکشن یونټ (ماډل 28193) لپاره جامع عملیاتي لارښوونې، د خوندیتوب لارښوونې، تخنیکي مشخصات، او د ساتنې پروسیجرونه چمتو کوي، چې یو مسلکي طبي سکشن وسیله ده.
مخکېview د ګیما ټوبي لارښود سکشن اسپریټر - د کارونکي لارښود او تخنیکي مشخصات
د GIMA TOBI MANUALE سکشن اسپریټر لپاره جامع کارونکي لارښود او تخنیکي مشخصات. د هغې د ځانګړتیاوو، خوندیتوب لارښوونو، پاکولو، ساتنې، او ستونزو حل کولو په اړه زده کړه وکړئ.
مخکېview GIMA S/S Bierer Aspiration Tube - User Manual & Specifications (Models 28262, 28264)
Official user manual and technical specifications for the GIMA S/S Bierer Aspiration Tube (Models 28262 and 28264). Details intended use, features, procedure, cleaning, sterilization, operating conditions, and manufacturer information.
مخکېview د GIMA پلس آکسیمیټر پروب کارونکي لارښود - تخنیکي مشخصات او د کارونې لارښود
د GIMA د نبض آکسیمیټر پروبونو لپاره جامع کارونکي لارښود، چې د محصول مشخصات، مطلوب کارول، کلینیکي ګټې، ساتنه، پاکول، غیر منتن کول، چاپیریالي اړتیاوې، او الکترو مقناطیسي مطابقت (EMC) معلومات پوښي. د ماډل شمیرې ESA0015، ESC0029، او نور شامل دي.
مخکېview GIMA KYOLING® Manual Suction Pump with Filter - Product Information
Comprehensive guide for the GIMA KYOLING® manual suction pump with filter, detailing its features, specifications, usage instructions, maintenance procedures, and troubleshooting tips for medical professionals.
مخکېview د GIMA فشار فعال شوي خوندیتوب لانسیټونه (PA) - د کیپلیري وینې S لپاره طبي وسیلهampلينګ
د GIMA فشار فعال شوي خوندیتوب لانسیټ (ماډل PA) په اړه معلومات، د کیپلیري وینې لپاره یو جراثیمي، یوځل کارول کیدونکی وسیلهampژبه. د کارولو اراده، لارښوونې، اخطارونه، ذخیره کول، او ضایع کول شامل دي.