GIMA 32775 UM-102A

GIMA 32775 UM-102A Mercury Free Sphygmomanometer User Manual

Model: 32775 UM-102A

پیژندنه

This manual provides essential information for the proper use, setup, operation, maintenance, and troubleshooting of the GIMA 32775 UM-102A Mercury Free Sphygmomanometer. Please read this manual thoroughly before using the device to ensure safe and effective operation.

The GIMA 32775 UM-102A is a professional, highly advanced digital blood pressure monitor designed as an alternative to traditional mercury column devices. It features auscultatory blood pressure measurement and a digital readout for pulse rate.

د خوندیتوب مهم معلومات

د محصول اجزا

The GIMA 32775 UM-102A Mercury Free Sphygmomanometer typically includes the following components:

GIMA 32775 UM-102A Mercury Free Sphygmomanometer main unit with cuff and bulb

شکل 1: GIMA 32775 UM-102A Mercury Free Sphygmomanometer, showing the main unit, cuff, and inflation bulb. The device is designed for desk use and features a digital display.

چمتو کول

د بسته بندۍ او ابتدايي تفتیش

  1. په احتیاط سره د بسته بندۍ څخه ټولې برخې لرې کړئ.
  2. Inspect the device and accessories for any signs of damage. Do not use the device if it appears damaged.
  3. ډاډ ترلاسه کړئ چې ټول لیست شوي اجزا شتون لري.

د وسیلې ځای په ځای کول

The GIMA 32775 UM-102A is a desk-type sphygmomanometer. Place the main unit on a stable, flat surface at a comfortable height for measurement.

شاته view of GIMA 32775 UM-102A showing cuff storage

شکل 2: شاته view of the GIMA 32775 UM-102A, illustrating the integrated cuff holder and grip for convenient storage and carrying.

د کف نښلول

  1. Locate the air tube connector on the main unit.
  2. Firmly insert the air tube from the cuff into the connector. Ensure a secure fit to prevent air leakage during measurement.

عملیاتي لارښوونې

د اندازه کولو دمخه

د کف پلي کول

  1. Place the cuff on your bare upper arm, approximately 2-3 cm above the elbow joint.
  2. Ensure the air tube runs down the center of your arm in line with your middle finger.
  3. Wrap the cuff snugly around your arm, ensuring it is not too tight or too loose. The cuff size is 32 cm.
  4. په آرامۍ سره کښیناستئ او خپلې پښې په فرش باندې کېږدئ، او لاس مو په میز باندې آرام کړئ ترڅو کف د زړه په کچه وي.
GIMA 32775 UM-102A Sphygmomanometer in use, showing arm placement

شکل 3: Proper arm and device placement for blood pressure measurement using the GIMA 32775 UM-102A. The image shows the device's height relative to an adult's arm.

Taking a Measurement (Auscultatory Method)

This device uses the auscultatory method, requiring a stethoscope (not included) for listening to Korotkoff sounds.

  1. Press the "ON/OFF" button to power on the device. The digital display will activate.
  2. Place the stethoscope diaphragm over the brachial artery in the antecubital fossa (inner elbow).
  3. Close the air release valve on the inflation bulb.
  4. Inflate the cuff rapidly by squeezing the inflation bulb until the pressure displayed on the digital column is approximately 20-30 mmHg above the expected systolic pressure, or until the radial pulse is no longer palpable.
  5. Slowly open the air release valve to deflate the cuff at a rate of 2-3 mmHg per second.
  6. Listen carefully through the stethoscope:
    • The pressure reading on the digital column when you first hear clear, tapping sounds (Korotkoff Phase I) is the د وینې سیسټولیک فشار.
    • Continue deflating. The pressure reading on the digital column when the sounds disappear completely (Korotkoff Phase V) is the ډیسټولیک د وینې فشار.
  7. Note the systolic and diastolic readings from the digital display. The device also provides a digital readout for pulse rate.
  8. Fully deflate the cuff and remove it from your arm.
  9. Press the "ON/OFF" button again to turn off the device, or it will automatically power off after a period of inactivity.

ساتنه او پالنه

پاکول

ذخیره کول

کیلیبریشن

It is recommended to have the sphygmomanometer calibrated periodically by qualified personnel to ensure accuracy. Consult your supplier or manufacturer for calibration services.

د ستونزو حل کول

ستونزهاحتمالي لاملد حل لاره
کله چې بریښنا فعاله وي نو هیڅ ښودنه نشته.Power source issue (if applicable, though this is manual).Ensure the device is properly activated. If battery-operated (not specified, but common for digital displays), check batteries.
کف نه پړسېږي یا ورو ورو پړسېږي.Air leakage from cuff or tubing; valve not closed.Check all connections for tightness. Ensure the air release valve is fully closed before inflation. Inspect cuff and tubing for damage.
ناسم لوستل.Improper cuff placement; patient movement; need for calibration.Re-apply cuff correctly. Remain still during measurement. Consider professional calibration.
د کوروتکوف غږونو اورېدلو کې ستونزه.Stethoscope placement; ambient noise; technique.Ensure correct stethoscope placement. Measure in a quiet environment. Practice auscultation technique.

If problems persist, contact customer support or a qualified service technician.

مشخصات

فیچرتفصیل
د ماډل شمیره32775 UM-102A
برانډGIMA
د اندازه کولو طریقهAuscultatory
د ښودلو ډولډیجیټل
د بریښنا سرچینهلارښود
شاملې برخېکف
د محصول ابعاد35 x 10 x 9 سانتي متره
د توکي وزن0.5 کیلوګرامه
د کف اندازه32 سانتي متره
د موادو ځانګړتیادوامدار
د عمر حد (تفصیل)بالغ
د اصلي هیوادایټالیا

د تضمین معلومات

Specific warranty details for the GIMA 32775 UM-102A Sphygmomanometer are typically provided at the point of purchase or within separate warranty documentation. Please retain your proof of purchase for warranty claims.

For information regarding the manufacturer's warranty, please refer to the documentation included with your product or contact GIMA directly.

د پیرودونکي ملاتړ

For technical assistance, service, or inquiries regarding your GIMA 32775 UM-102A Mercury Free Sphygmomanometer, please contact your local distributor or the manufacturer:

جوړونکی: GIMA

Refer to the official GIMA webد اړیکو د تازه معلوماتو لپاره سایټ یا د محصول بسته بندي.

اړوند اسناد - 32775 UM-102A

مخکېview د GIMA مادرید انیرایډ سفیګمومانومیټر - د مسلکي طبي محصول معلومات
د GIMA مادرید انیرایډ سفیګمومانومیټر (ماډل HS-201Q3) لپاره جامع لارښود، د هغې ځانګړتیاوې، د کارولو لارښوونې، مشخصات، او ساتنه په تفصیل سره بیانوي. دا سند د روغتیا پاملرنې مسلکیانو او کاروونکو لپاره اړین معلومات چمتو کوي.
مخکېview د GIMA ARM-30E+ ډیجیټل د وینې فشار مانیټر کارونکي لارښود
د GIMA ARM-30E+ ډیجیټل د وینې فشار مانیټر لپاره د کارونکي لارښود، د کارولو، خوندیتوب احتیاطاتو، ساتنې، او ستونزو حل کولو لارښوونې چمتو کوي.
مخکېview د GIMA KD-735 د لاس د وینې فشار مانیټر کارونکي لارښود
د GIMA KD-735 الکترونیکي لاس د وینې فشار مانیټر لپاره د کارونکي لارښود. د کور د وینې فشار دقیق څارنې لپاره د تنظیم، عملیاتو، خوندیتوب، ستونزو حل کولو او مشخصاتو په اړه لارښوونې چمتو کوي.
مخکېview د GIMA KD-7920 د لاس د وینې فشار د څارنې عملیاتي لارښود
د GIMA KD-7920 د لاس د وینې فشار مانیټر لپاره رسمي عملیات او ساتنې کتاب، د تنظیم، کارولو، ستونزو حل کولو، او مشخصاتو توضیحات.
مخکېview د ګیما KD-735 لاس د وینې فشار اتوماتیک نظارت کارونکي لارښود
د ګیما KD-735 د لاس اتوماتیک د وینې فشار مانیټر لپاره جامع عملیاتي لارښود، د کور او مسلکي روغتیا دقیق څارنې لپاره د تنظیم، کارولو، ستونزو حل کولو او ساتنې توضیحات.
مخکېview د GIMA KD-558 د لاس د وینې فشار مانیټر: د کارونکي لارښود او مشخصات
د GIMA KD-558 الکترونیکي سفیګمومانومیټر (د لاس د وینې فشار مانیټر) لپاره جامع کارونکي لارښود او تخنیکي مشخصات. د تنظیم، عملیاتو، خوندیتوب، ستونزو حل کولو او ساتنې په اړه زده کړه وکړئ.